Abstract:
Techniques are described which allow the traceable measurement, at low uncertainty levels, of changes in attenuation value in excess of 140 dB at frequencies from 0.5 MHz...Show MoreMetadata
Abstract:
Techniques are described which allow the traceable measurement, at low uncertainty levels, of changes in attenuation value in excess of 140 dB at frequencies from 0.5 MHz to 100 MHz. The coherent nature of the dual channel techniques employed allows the changes in the path length through an attenuator, produced when its setting is changed, to be measured accurately even at high attenuation levels.<>
Published in: IEEE Transactions on Instrumentation and Measurement ( Volume: 44, Issue: 2, April 1995)
DOI: 10.1109/19.377838