Abstract:
This paper describes the development of a Non-invasive automatic blood pressure monitor (BPM) based on a capacitive pressure sensor (CPS) with the enhanced capture, compa...Show MoreMetadata
Abstract:
This paper describes the development of a Non-invasive automatic blood pressure monitor (BPM) based on a capacitive pressure sensor (CPS) with the enhanced capture, compare, pulse width modulation (PWM) module (ECCP) of the microcontroller (MCU) and Bluetooth radio communications for home health monitoring. The objective of this paper is to develop automatic blood pressure device with accuracy and low cost by using a capacitive pressure sensor and logic inverter oscillator instead of the instrumentation amplifier and bandpass filter. The biomedical real-time blood pressure (BP) signal is measured by using pressure to capacitance circuit with pulse width modulation (PWM) technology and interfaced with PIC18F4550 microcontroller. The microcontroller detects the blood pressure signal and by using enhanced module functions, with the help of proposed algorithm, the blood pressure result (systolic, diastolic, pulse) has been calculated and transmitted to the Android smartphone health application via Bluetooth communication with the presence of mini LCD display, the records are stored in the internal memory of the microcontroller. The obtained result of the proposed device was verified with standard Sphygmomanometer results.
Published in: 2018 International Conference on Engineering Technology and their Applications (IICETA)
Date of Conference: 08-09 May 2018
Date Added to IEEE Xplore: 11 September 2018
ISBN Information: