Loading [MathJax]/extensions/MathMenu.js
Auto-phase calibration loop of a transmission array for focused microwave thermotherapy | IEEE Conference Publication | IEEE Xplore

Auto-phase calibration loop of a transmission array for focused microwave thermotherapy


Abstract:

In this paper, we introduce an automatic phase calibration loop circuit for a focused microwave thermotherapy system for those with diseases of the leg. An extended time ...Show More

Abstract:

In this paper, we introduce an automatic phase calibration loop circuit for a focused microwave thermotherapy system for those with diseases of the leg. An extended time reversal method is also introduced. The loop circuit is evaluated by measuring the phase errors at the array antenna feeding port. These measurements are compared to those without a loop. The results show that the overall phase errors are reduced enough to focus fields at a target point.
Date of Conference: 09-14 July 2017
Date Added to IEEE Xplore: 19 October 2017
ISBN Information:
Electronic ISSN: 1947-1491
Conference Location: San Diego, CA, USA

Contact IEEE to Subscribe

References

References is not available for this document.