Introduction
With the development of modern industry, there exists a trend of miniaturization of industrial devices. The study and application of gas-liquid two-phase flow in the millimeter-scale pipeline have become more and more important [1]–[3]. Bubble flow and slug flow are common phenomena of gas-liquid two-phase flow in millimeter-scale pipeline. The online measurement of the bubble/slug velocity is of great importance for academic research and industrial application [1]–[3]. However, with the decrease of pipe scale, the parameter measurement becomes more and more difficult, the conventional bubble/slug velocity measurement methods cannot meet the higher and higher requirement of academic research and industrial application of gas-liquid two-phase flow in millimeter-scale pipeline [1]–[5].
Combing the conductance sensors and the cross-correlation velocity measurement technique to implement the velocity measurement is a classic method and has been widely studied and applied by scientists and engineers for many decades in the research field of gas-liquid two-phase flow [1], [6]–[13]. Many achievements and technical progresses have been obtained [1], [6]–[13]. However, the conventional conductance sensors still have two disadvantages: 1) Most of the sensors are developed for the gas-liquid two-phase flow in normal-scale pipelines, the studies and applications on the bubble/slug velocity measurement for the gas-liquid two-phase flow in millimeter-scale pipelines are relatively limited [1], [6]–[13]. 2) Most of the sensors are developed on the basis of contact conductivity measurement technique, the electrodes are contacting directly with measured fluid, which will cause polarization effect and electrochemical erosion effect. Meanwhile, if the electrodes are contaminated, unpredictable measurement errors may arise [1], [6]–[17].
Capacitively coupled contactless conductivity detection (C4D) technique is a contactless conductivity measurement technique and its electrodes are not directly in contact with the measured fluid. With this kind of conductivity measurement technique, the disadvantages of the conventional contact conductance sensors (polarization effect and electrochemical erosion effect) can be avoided [14]–[23]. However, as a developing technique, C4D is mainly studied and applied in the research field of analytical chemistry for measuring ion concentration or solution conductivity in capillaries (the inner diameter of capillary is usually less than 0.2 mm). Up to date, research works on the application of C4D in two-phase flow research field are very limited [14]–[23].
This work aims to propose a new bubble/slug velocity measurement method by combining C4D technique and cross-correlation velocity measurement technique. Firstly, a new C4D sensor which is suitable for bubble/slug velocity measurement in millimeter-scale pipelines is developed. In this new sensor, series resonance principle is introduced to overcome the unfavorable influence of coupling capacitances and simulated inductor technique is introduced to overcome the disadvantages of a practical inductor. Secondly, based on the cross-correlation velocity measurement technique, a bubble/slug velocity measurement system is developed. Finally, with two conductance signals obtained by two C4D sensors (the upstream sensor and the downstream sensor), the bubble/slug velocity is determined by the cross-correlation velocity measurement technique.
Design of the New C4D Sensor
A. Principle of C4D Sensor
Fig. 1(a) shows the construction of a typical radial C4D sensor. Fig. 1(b) is the simplified equivalent circuit of the C4D sensor.
Principle of C4D sensor: (a) The construction of a typical radial C4D sensor, (b) Simplified equivalent circuit of C4D sensor.
B. New C4D Sensor
As shown in Fig. 1, two coupling capacitances
Based on our previous research works on C4D technique, the unfavorable influence of the coupling capacitances
Meanwhile, our previous research results also indicate that the simulated inductor technique is an effective way to provide the inductive reactance and a simulated inductor can overcome the disadvantages of a practical inductor (such as non-adjustable inductance value, large internal resistance and difficulties to implement large-valued inductor) [24]. So, in this work, the simulated inductor technique is adopted.
Fig. 2. illustrates the circuit of the new simulated inductor used in this work. The new simulated inductor is developed on the basis of the classic Riordan simulated inductor [25]–[29].
The equivalent inductance \begin{align} L_{eq}=&\frac {R_{1} R_{2} (R_{3} +R_{4} )}{R_{5} }C_{m} \\ R_{eq}=&\frac {R_{1} R_{2} (R_{3} +R_{4} )}{R_{5} R_{6} } \end{align}
According to Equation (1), the value of the equivalent inductance can be determined by
Fig. 3 shows the construction of the new C4D sensor which can be mainly divided into two parts, the detection path and the signal processing unit. The detection path consists of an AC source, an excitation electrode, an insulating pipeline, a detection electrode and the new simulated inductor. The signal processing unit consists of a current voltage conversion circuit (I/V circuit), a full wave rectifier, a low pass filter and a data acquisition module.
Fig. 4 shows the simplified equivalent circuit of the detection path of the new C4D sensor. The impedance of the detection path \begin{equation} Z=R_{x} +R_{eq} +j\left({2\pi fL_{eq} -\frac {C_{1} +C_{2} }{2\pi fC_{1} C_{2} }}\right) \end{equation}
At series resonance:\begin{equation} 2\pi f_{0} L_{eq} -\frac {C_{1} +C_{2} }{2\pi f_{0} C_{1} C_{2} }=0 \end{equation}
The introduction of the inductor \begin{equation} f=f_{0} =\frac {1}{2\pi }\sqrt {\frac {C_{1} +C_{2} }{L_{eq} C_{1} C_{2} }} \end{equation}
\begin{equation} Z=R_{x} +R_{eq} \end{equation}
\begin{equation} I_{o} =\frac {U_{i} }{R_{x} +R_{eq} } \end{equation}
After the operation of I/V circuit the output current \begin{equation} U_{o} =-I_{o} R_{f} =-\frac {R_{f} }{R_{x} +R_{eq} }U_{i} \end{equation}
Bubble/Slug Velocity Measurement
Fig. 5 shows the construction of the measurement system for the bubble/slug velocity measurement.
The bubble/slug velocity measurement is implemented by the cross-correlation velocity measurement technique.
Two same new C4D sensors are installed on the upstream and the downstream of the pipeline respectively to obtain two independent signals (the upstream signal \begin{equation} R_{xy} (i)=\frac {1}{N}\sum \limits _{n=1}^{N} {x(n\Delta t)} y[(n+i)\Delta t] \end{equation}
By searching the time position of the maximum cross-correlation coefficient, the transit time
Finally, according to the cross-correlation velocity measurement principle [8], [9], the bubble /slug velocity \begin{equation} v=k\frac {L}{\tau } \end{equation}
Where,
Experimental Setup
Fig. 6 illustrates the experimental setup for the bubble/slug velocity measurement.
The velocity measurement experiments are carried out in three horizontal glass pipes with inner diameters of 4.50 mm, 5.46 mm and 6.44 mm respectively. Correspondingly, three pairs of C4D sensors are developed. Table I lists the detailed information of the C4D sensors used in the velocity measurement experiments. The AC source is the signal generator RIGOL DG 1022. The data acquisition module is the NI cDAQ-9172. The bubble velocity measurement of bubble flow and the slug velocity measurement of slug flow in millimeter-scale pipelines are investigated. Fig. 7(a) is a photo of typical bubble flow in 4.50 mm pipeline, Fig. 7(b) is a photo of typical slug flow in 4.50 mm pipeline; Fig. 8(a) is the measured signal without bubble/slug in 4.50 mm pipeline, Fig. 8(b) is the measured signal with bubble in 4.50 mm pipeline, Fig. 8(c) is the measured signal with slug in 4.50 mm pipeline.
(a) Measured signal without bubble/slug; (b) Measured signal with bubble; (c) Measured signal with slug.
The experimental materials are tap water (the conductivity of the tap water ranges from 0.17~0.19 mS/cm) and nitrogen. The ranges of the equivalent resistors of the fluid between the two electrodes
Tap water and nitrogen are driven into the pipeline by high-pressure nitrogen. The liquid flowrate is measured by an electromagnetic flowmeter (IFC-300C, Krohne) and the gas flowrate is measured by a thermal gas flowmeter (F-201 CB, Bronkhorst). By adjusting the different flowrate combinations, different flow patterns and different bubble/slug velocities can appear in the pipelines. In this work, the bubble velocity ranges from 0.09 m/s to 0.53 m/s and the slug velocity ranges from 0.06 m/s to 1.08 m/s.
The reference velocity data are obtained by a high-speed camera (MotionXtraN-4, IDT Red-lake) and a 60 mm f/2.8 D lens (AF Micro-Nikkor). From the captured pictures, the reference velocity of bubble/slug can be calculated by:\begin{equation} v_{r} =\frac {L_{c} }{nT_{c} } \end{equation}
Experimental Results and Discussion
Fig. 9. to Fig. 11. show the bubble and slug velocity measurement results. The relative error \begin{equation} e=\frac {v-v_{r} }{v_{r} }\times 100\% \end{equation}
Experimental results of 5.46 mm pipe (
Experimental results of 6.44 mm pipe (
Table III lists the maximum relative errors and the velocity ranges of the bubble velocity measurement and the slug velocity measurement.
These experimental results show that the proposed method for bubble/slug velocity measurement in millimeter-scale pipelines is effective. The new C4D sensor, which is developed on the basis of series resonance principle and simulated inductor technique, is successful. And, the measurement accuracy of the bubble/slug velocity measurement system is satisfactory. The maximum relative error of bubble flow velocity measurement is 5.41%, and the maximum relative error of slug flow velocity measurement is 4.90%. The maximum relative error appears in 6.44 mm pipe. 6.44 mm is a little over the range of the critical capillary size of the pipe for ‘mini’ (or ‘micro’) scale, and in this pipe, the effect of superficial tension is not high enough to ensure bubble/slug stable. Maybe, that is the reason of higher error in 6.44 mm pipe.
Besides, it is necessary to indicate that the proposed method still has some drawbacks. When the gas-liquid two-phase flow is the bubble flow with smaller bubbles and lower velocity (as shown in Fig. 12.), the accuracy of bubble velocity measurement may decrease (i.e., greater measurement error may arise) because the obtained signals is too weak to realize a successful cross-correlation velocity measurement in this situation. To further improve the measurement performance of the proposed method (and the developed new C4D sensor) would be our research work in the future.
The typical photo of bubble flow with low velocity and small bubble diameter in 4.50 mm pipeline.
Conclusion
This work focuses on the bubble/slug velocity measurement of gas-liquid two-phase flow in millimeter-scale pipelines. A new bubble/slug velocity measurement method is proposed by combining C4D technique and cross-correlation velocity measurement technique. With the introductions of series resonance principle and simulated inductor technique, a new C4D sensor which is suitable for bubble/slug velocity measurement in millimeter-scale pipelines is developed. In this new C4D sensor, the simulated inductor technique is introduced to develop an inductor module and hence to overcome the unfavorable influence of the coupling capacitances by series resonance principle. A bubble/slug velocity measurement system is constructed. Two independent signals are acquired by two new C4D sensors (the upstream sensor and the downstream sensor) and the bubble/slug velocity measurement is realized by the cross-correlation velocity measurement technique.
Experiments are carried out to verify the effectiveness of the proposed method. The experimental results indicate that the new contactless method for bubble/slug velocity measurement in millimeter-scale pipelines is effective. The development of new C4D sensor is successful. The bubble/slug velocity measurement system can realize the online and contactless velocity measurement successfully and the measurement accuracy of the bubble/slug velocity is satisfactory. The maximum relative error of bubble flow velocity measurement is 5.41%, and the maximum relative error of slug flow velocity measurement is 4.90%.
Useful knowledge and experience are obtained. The study results can provide a good reference for others’ research works. However, due to the complexity of gas-liquid two-phase flow in millimeter-scale pipeline, more research works should be undertaken in this area.