Abstract:
Contactless conductivity detection has attained great attention in the last decade in capillary electrophoresis. But contactless conductivity detection is restricted by i...Show MoreMetadata
Abstract:
Contactless conductivity detection has attained great attention in the last decade in capillary electrophoresis. But contactless conductivity detection is restricted by its sensitivity, how to improve detection sensitivity has become the key of contactless conductivity. In this paper a novel extended model for the detection cell consisting of a network of resistors and capacitors is proposed according to the structure of contactless conductivity detection. The effect of the detector geometry on the sensitivity of contactless conductivity is studied. By simulation, the optimal parameters are obtained. Sandwich electrode structure was established to improve the performance of the detector and minimize the stray capacitance between the electrodes. Using the optimal electrode structure, the two peaks corresponding to K+ and Mg2+ are clearly resolved with complete separation.
Published in: The 8th Annual IEEE International Conference on Nano/Micro Engineered and Molecular Systems
Date of Conference: 07-10 April 2013
Date Added to IEEE Xplore: 18 July 2013
ISBN Information: