It is pointed out that to restore experimental computed tomography (CT) images which have been blurred by a spatially variant point spread function (PSF), a quadrant symm...Show MoreMetadata
Abstract:
It is pointed out that to restore experimental computed tomography (CT) images which have been blurred by a spatially variant point spread function (PSF), a quadrant symmetry method which simultaneously optimizes storage requirements for the estimated PSFs and computational speed is used. The quadrant symmetry approach required less than 0.579 Megawords of storage for 9*9 pixel spatially variant PSFs in a 256*256 pixel image, allowing image restoration in Cartesian coordinates. A locally adaptive restoration method formulated using a noise prewhitening filter derived from the measured noise power spectrum, the conjugate gradient method was used to obtain a numerical solution. Restored images for a resolution phantom and for a cadaver femur are presented.<>
Published in: IEEE Transactions on Medical Imaging ( Volume: 11, Issue: 4, December 1992)
DOI: 10.1109/42.192690