Abstract:
In this paper, we analyze the influence of the superconducting quantum interference device (SQUID) gradiometer baseline on the recording of magnetocardiographic measureme...Show MoreMetadata
Abstract:
In this paper, we analyze the influence of the superconducting quantum interference device (SQUID) gradiometer baseline on the recording of magnetocardiographic measurements. The magnetometers consist of high-temperature superconducting radio-frequency SQUIDs fabricated from YBaCuO thin films, and a substrate resonator which serves as tank circuit. The gradiometers are formed using two or three such magnetometers with individual readouts in electronic difference. We have compared the measurement results using a magnetometer and first- and second-order gradiometers with different baselines. In a standard magnetically shielded room, we found not only an increasing signal-to-noise ratio in adult magnetocardiographic measurements, but also a decreasing distortion of the magnetic field map with increasing baseline of the gradiometer. Using a first-order gradiometer with an ultralong baseline of 18 cm, we have successfully measured the heart signal of a fetus in real time.
Published in: IEEE Transactions on Applied Superconductivity ( Volume: 13, Issue: 4, December 2003)