Abstract:
A new generation of implantable, telemetric transmitters for intracranial pressure (ICP) measurements have been developed. A unique technique used in packaging the silico...Show MoreMetadata
Abstract:
A new generation of implantable, telemetric transmitters for intracranial pressure (ICP) measurements have been developed. A unique technique used in packaging the silicon piezoresistive pt essure transducer provides excellent long-term stability. Pulse code modulation is used for data transmission over a radio frequency (RF) link. To minimize the component count, two semicustom, bipolar integrated circuits are used. The transmitter electronics are housed inside a 29 ×20 ×7 mm titanium package along with the pressure transducer and two lithium batteries. Even though the transmitter consumes less than 0.4 mW of power, it is turned on remotely via RF signal transduction only on demand in order to extend the lifetime of the batteries to years. The pressure input of the transmitter has a dynamic range of -100- +200 mmHg with a 0.3 mmHg resolution and a 1 mmHg accuracy. Long-term in vitro and in vivo pressure baseline stabilities of better than 1 and 2 mmHg per month, respectively, have been achieved.
Published in: IEEE Transactions on Biomedical Engineering ( Volume: BME-33, Issue: 4, April 1986)