Abstract:
Dissipation factor (DF) standards obtained by connecting a shielded three-terminal capacitor in series with a shielded resistor have been developed for calibration purpos...Show MoreMetadata
Abstract:
Dissipation factor (DF) standards obtained by connecting a shielded three-terminal capacitor in series with a shielded resistor have been developed for calibration purposes. An analysis of these DF standards, including precautions in their construction and use, is presented and calibration procedures using the NIST high voltage capacitance bridge is discussed.
Date of Conference: 06-10 July 1998
Date Added to IEEE Xplore: 06 August 2002
Print ISBN:0-7803-5018-9